Loughborough University
Browse
DDTR_Cheung_et_al_2015.pdf (2.02 MB)

Delivery of large molecular protein using flat and short microneedles prepared using focused ion beam (FIB) as a skin ablation tool

Download (2.02 MB)
journal contribution
posted on 2015-09-15, 15:22 authored by Karmen Cheung, Geoff West, Diganta DasDiganta Das
Many studies have been reported in the literature on the effects of various geometries and lengths of microneedles (MNs) on transdermal drug delivery using a variety of drug molecules. In particular, sharp-tipped MNs have been used to disrupt the top layer of the skin, namely, stratum corneum (SC). It has also been shown that short- and flat-tipped MNs can pierce the SC and they have the potential to increase drug permeability. However, there is little work that explores MNs as a skin ablative tool with a view to increasing skin permeability. To address this point, well-defined small patterns (size of individual pattern 10–20 μm) on the tip of flat MN (tip radius of individual MN ∼250 μm) were created and their effects evaluated on the permeability of bovine serum albumin (BSA), which is chosen as a model drug of high molecular weight. The patterns on the tip of flat MN act as rough surfaces (e.g. like sand paper) which when applied on the surface of the skin ablate the SC layer. Focused ion beam (FIB) has been used as the fabrication technique for the MNs. The permeability data are then compared with the other data for flat- and sharp-tipped MN. The permeability data from passive diffusion experiments are used as the reference case. The exact number of MNs or patterns in the flat and patterned MN patches is not considered as important as they have not been designed to pierce the skin. However, this is an important consideration in the case of sharp MNs as they pierce and create cavities in the skin. It is found that the delivery of BSA with the fabricated flat and patterned MNs gave similar but somewhat lower drug permeation profile in comparison to the sharp MNs. Passive diffusion showed no permeation, as would be expected due to the large size of the chosen molecule.

Funding

This work was supported by Loughborough University, UK, and the EPSRC, UK.

History

School

  • Aeronautical, Automotive, Chemical and Materials Engineering

Department

  • Chemical Engineering

Published in

Drug Delivery and Translational Research

Volume

5

Issue

4

Pages

462 - 467

Citation

CHEUNG, K., WEST, G. and DAS, D., 2015. Delivery of large molecular protein using flat and short microneedles prepared using focused ion beam (FIB) as a skin ablation tool. Drug Delivery and Translational Research, 5 (4), pp.462-467.

Publisher

Springer US (© the authors)

Version

  • VoR (Version of Record)

Publisher statement

This work is made available according to the conditions of the Creative Commons Attribution 4.0 International (CC BY 4.0) licence. Full details of this licence are available at: https://creativecommons.org/licenses/by/4.0/

Publication date

2015

Notes

This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http:// creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

ISSN

2190-393X

eISSN

2190-3948

Language

  • en