Title: Medical rapid prototyping technologies: state of the art and current limitations for application in oral and maxillofacial surgery
Authors: Winder, John
Bibb, Richard J.
Keywords: Medical rapid prototyping
Computed tomography
Stereolithography
Fused deposition modelling
Artefacts
Issue Date: 2005
Publisher: © Elsevier
Citation: WINDER, J. and BIBB, R.J., 2005. Medical rapid prototyping technologies: state of the art and current limitations for application in oral and maxillofacial surgery. Journal of Oral and Maxillofacial Surgery, 63 (7), pp.1006-1015.
Abstract: Purpose We describe state of the art software and hardware requirements for the manufacture of high quality medical models manufactured using medical rapid prototyping. The source of the medical model artefacts and there physical appearance are illustrated along with remedies for their removal. Materials and Methods Medical models were built using predominantly stereolithography and fused deposition modelling at both institutions over a period of 6 years. A combined total of 350 models have been produced for a range of maxillofacial, neurosurgical and orthopaedic applications. Stereolithography, fused deposition modelling computerised numerical milling and other technologies are described. Results A range of unwanted artefacts that create distortions on medical models have been identified. These include, data import, CT gantry distortion, metal, motion, surface roughness due to support structure removal or surface modelling and image data thresholding. The source of the artefact has been related to the patient, imaging modality performance or the modelling technology. Discussion as to the significance of the artefacts on clinical use is provided. Conclusions It is recommended that models of human anatomy generated by medical rapid prototyping are subject to rigorous quality assurance at all stages of the manufacturing process. Clinicians should be aware of potential areas for inaccuracies within models and review the source images in cases where model integrity is in doubt.
Description: This is the author’s version of a work that was accepted for publication in Journal of Oral and Maxillofacial Surgery. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version was subsequently published at: http://dx.doi.org/10.1016/j.joms.2005.03.016
Version: Accepted for publication
DOI: 10.1016/j.joms.2005.03.016
URI: https://dspace.lboro.ac.uk/2134/14449
Publisher Link: http://dx.doi.org/10.1016/j.joms.2005.03.016
ISSN: 0278-2391
Appears in Collections:Published Articles (Design School)

Files associated with this item:

File Description SizeFormat
Winder-Bibb-Medical-RP-JOMS-2005.pdfAccepted version91.17 kBAdobe PDFView/Open

 

SFX Query

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.