Title: Haematoma detection using EIT in a sheep model
Authors: Ayati, S. Bentolhoda
Bouazza-Marouf, Kaddour
Kerr, David
O'Toole, Michael D.
Keywords: Electrical impedance tomography
Sheep model
Assistive medical technology
Medical image processing
Issue Date: 2014
Publisher: © ACTA Press
Citation: AYATI, S.B. ... et al, 2014. Haematoma detection using EIT in a sheep model. International Journal of Computers and Applications, 36 (3), 7pp.
Abstract: Performance evaluation of a portable digital electrical impedance tomography system to detect haematomas using phantom and sheep models is presented. Experiments have been performed using 8-electrode full array configuration. Conductivity changes were detected in phantom model while anomalies were placed at center and close to the edge of the tank. Bleeding rate was successfully monitored in sheep model while blood-like conductivity solution was injecting via the brainstem. EIT images were reconstructed sequentially for different injection volumes and the quantity index (QI) was calculated as a function of the injected solution volume. The results show a linear relationship of QI to the injected volume. Images of the sheep experiment with the simulated haematomas, blood-like conductivity gel, placed on top of the parietal lobes of the brain on the left and right sides were reconstructed and haematomas are clearly detected and localized. These experiments prove that the detection and quantification of haematomas in the brain is possible and encourage further investigation for medical applications.
Description: This paper was submitted for publication in the journal, International Journal of Computers and Applications. The definitive version is available at: http://dx.doi.org/10.2316/Journal.202.2014.3.202-3834
Version: Submitted for publication
DOI: 10.2316/Journal.202.2014.3.202-3834
URI: https://dspace.lboro.ac.uk/2134/16548
Publisher Link: http://dx.doi.org/10.2316/Journal.202.2014.3.202-3834
ISSN: 1206-212X
Appears in Collections:Published Articles (Mechanical, Electrical and Manufacturing Engineering)

Files associated with this item:

File Description SizeFormat
202-3834 HAEMATOMA DETECTION USING EIT in a SHEEP MODEL.pdfSubmitted version341.15 kBAdobe PDFView/Open


SFX Query

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.