Title: Computational evaluation of artery damage in stent deployment
Authors: He, Ran
Zhao, Liguo
Silberschmidt, Vadim V.
Liu, Yang
Vogt, Felix
Keywords: Artery damage
Hyperelastic damage model
Finite element
Pre/post-dilation
Stent deployment
Issue Date: 2018
Publisher: Elsevier BV © The Authors
Citation: HE, R. ... et al, 2018. Computational evaluation of artery damage in stent deployment. Procedia Structural Integrity, 13, pp.187-191.
Abstract: This paper aims to evaluate damage in an arterial wall and plaque caused by percutaneous coronary intervention using a finite-element (FE) method. Hyperelastic damage models, verified against experimental results, were used to describe stress-stretch responses of arterial layers and plaque in the lumen; these models are capable to simulate softening behaviour of the tissue due to damage. Abaqus CAE was employed to create the FE models for an artery wall with two constituent layers (media and adventitia), a symmetric uniform plaque, a bioresorbable polymeric stent and a tri-folded expansion balloon. The effect of percutaneous coronary intervention on vessel damage was investigated by simulating the processes of vessel pre-dilation, stent deployment and post-stenting dilation. Energy dissipation density was used to assess the extent of damage in the tissue. Overall, the plaque experienced the most severe damage due to its direct contact with the stent, followed by the media and adventitia layers. Softening of the plaque and the artery due to the pre-dilation-induced damage can facilitate the subsequent stent-deployment process. The plaque and artery experienced heterogeneous damage behaviour after the stent deployment, caused by non-uniform deformation. The post-stenting dilation was effective to achieve a full expansion of the stent but caused additional damage to the artery. The computational evaluation of artery damage can be also potentially used to assess the risk of in-stent restenosis after percutaneous coronary intervention.
Description: This is an Open Access article. It is published by Elsevier under the Creative Commons Attribution 4.0 International Licence (CC BY). Full details of this licence are available at: http://creativecommons.org/licenses/by/4.0/. This is part of special issue: ECF22 - Loading and Environmental effects on Structural Integrity edited by Aleksandar Sedmak, Zoran Radaković and Marko Rakin.
Sponsor: British Heart Foundation (Grant number: FS/15/21/31424; Title: Towards controlling the mechanical performance of polymeric bioresorbable vascular scaffold during biodegradation) and the UK Royal Society (Grant number: IE160066; Title: Evaluating the Performance of Additively Manufactured Endovascular Scaffolds).
Version: Published
DOI: 10.1016/j.prostr.2018.12.031
URI: https://dspace.lboro.ac.uk/2134/37608
Publisher Link: https://doi.org/10.1016/j.prostr.2018.12.031
ISSN: 2452-3216
Appears in Collections:Conference Papers and Presentations (Mechanical, Electrical and Manufacturing Engineering)

Files associated with this item:

File Description SizeFormat
Ran conference paper.pdfPublished version748.71 kBAdobe PDFView/Open

 

SFX Query

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.